Error limits for thermometers
The following error limits are according to "Eichordnung EO 14-1", the German Federal Weights and Measures Regulations.
For thermometers calibrated for total immersion containing non-wetting thermometric liquid (e.g., mercury and mercury-thallium alloy) and for thermometers containing wetting thermometric liquid (e.g., toluene, pentane and petroleum).
Official error limits for thermometers containing non-wetting thermometric liquids, for the subdivisions:
| Temperature range from °C / to °C |
0.05 °C | 0.1 °C | 0.2 °C | 0.5 °C | 1 °C | 2 °C | 5 °C |
| -58 / -10 | - | ± 0.3 | ± 0.4 | ± 0.5 | ± 1 | ± 2 | ± 5 |
| -10 / 110 | ± 0.1 | ± 0.2 | ± 0.3 | ± 0.5 | ± 1 | ± 2 | ± 5 |
| 110 / 210 | - | - | ± 0.4 | ± 0.5 | ± 1 | ± 2 | ± 5 |
| 210 / 410 | - | - | - | ± 1 | ± 2 | ± 2 | ± 5 |
| 410 / 610 | - | - | - | - | ± 3 | ± 4 | ± 5 |
Official error limits for thermometers containing wetting thermometric liquids, for the subdivisions:
|
Temperature range from °C / to °C |
0.5 °C | 1 °C | 2 °C | 5 °C | |||
| -200 / -110 | - | ± 3 | ± 4 | ± 5 | |||
| -110 / -10 | ± 1 | ± 2 | ± 4 | ± 5 | |||
| -10 / 110 | ± 1 | ± 2 | ± 3 | ± 5 | |||
| 110 / 210 | - | ± 3 | ± 4 | ± 5 |
Calibration / Temperature correction (approximation)
Unless marked otherwise, thermometers are calibrated "for total immersion". Hence that the meniscus of the mercury column is level with the surface of the liquid being measured. If part of the mercury column is visible above the liquid surface, a correction may be necessary. Equation for correction: tk = t + (t-t')n/6250 |
|
Example: Mean temp. of emerging Length of emerging Corrected temp.: tk = 161.5 °C |