Uniform lid thickness ensures trouble-free piercing.
Consistent wall thickness.
Tight fitting attached lid provides leak-free seal, yet reopens easily.
High clarity.
Autoclavable at 121 °C (2 bar), acc. DIN EN 285.
Relative centrifugal force (RCF)
The quoted permissible RCF values (identical to g values) refer to filling with water and 20 min centrifugation time at 20°C. However, the actual stress limit is affected to a large extent by conditions such as the positioning in the rotor, chemicals, temperature and running time. There- fore, the RCF values must only be understood as a general guide. (DIN 58970).
Equation to calculate RCF: RCF = 1,118 x r x (n/1000)2
Example: Rotation radius r = 180 mm (vertical distance between center of rotation axis and bottom end of centrifuge tube) Speed n = 6000 min-1 (rotations per minute divided by 1000)