Comparison of error limits
| Bottle-top burette Titrette® | Bottle-top burette acc. to DIN EN ISO 8655-3 |
Glass burettes Class A acc. DIN EN ISO 385 |
||||||||
| Volume ml |
Partial volume ml |
A* | CV* | A* | CV* | EL** | ||||
| ≤± % | ≤± µl | ≤ % | ≤ µl | ≤± % | ≤± µl | ≤ % | ≤ µl | ± µl | ||
| 10 | 10 | 0.10 | 10 | 0.05 | 5 | 0.3 | 30 | 0.1 | 10 | 20 |
| 5 | 0.20 | 10 | 0.10 | 5 | 0.6 | 30 | 0.2 | 10 | 20 | |
| 1 | 1.00 | 10 | 0.5. | 5 | 3 | 30 | 1 | 10 | 20 | |
| 25 | 25 | 0.07 | 18 | 0.025 | 6 | 0.2 | 50 | 0.1 | 25 | 30 |
| 12.5 | 0.14 | 18 | 0.05 | 6 | 0.4 | 50 | 0.2 | 25 | 30 | |
| 2.5 | 0.70 | 18 | 0.25 | 6 | 2 | 50 | 1 | 25 | 30 | |
| 50 | 50 | 0.06 | 30 | 0.02 | 10 | 0.2 | 100 | 0.1 | 50 | 50 |
| 25 | 0.12 | 30 | 0.04 | 10 | 0.4 | 100 | 0.2 | 50 | 50 | |
| 5 | 0.60 | 30 | 0.20 | 10 | 2 | 100 | 1 | 50 | 50 | |
* A = accuracy. CV =
coefficient of variation
** EL = error limit (EL = A + 2CV)
The maximum resolution of the display:
10 ml instrument: 0.001 ml
25 ml instrument: 0.001 ml, above 20 ml titration volume 0.01 ml
50 ml instrument: 0.002 ml, above 20 ml titration volume 0.01 ml