Titrette®
Technical Data

Comparison of error limits

Fehlergrenzen
    Bottle-top burette Titrette® Bottle-top burette acc.
to DIN EN ISO 8655-3
Glass burettes Class A acc.
to DIN EN ISO 385
Volume
ml
Partial vol.
ml
A* CV* A* CV* EL**
≤± % ≤± µl ≤ % ≤ µl ≤± % ≤± µl ≤ % ≤ µl ± µl
25 25 0.07 18 0.025 6 0.2 50 0.1 25 30
12.5 0.14 18 0.05 6 0.4 50 0.2 25 30
2.5 0.70 18 0.25 6 2 50 1 25 30
50 50 0.06 30 0.02 10 0.2 100 0.1 50 50
25 0.12 30 0.04 10 0.4 100 0.2 50 50
5 0.60 30 0.20 10 2 100 1 50 50

*   A  = accuracy, CV = coefficient of variation
** EL = error limit (EL = A + 2CV)

The maximum resolution of the display:
25 ml instrument: 0.001 ml, above 20 ml titration volume 0.01 ml;
50 ml instrument: 0.002 ml, above 20 ml titration volume 0.01 ml.

Range of application

Fehlergrenzen
The instrument can be used for the following titration media (maximum concentration 1 mol/l):
Acetic acid Nitric acid Sodium arsenite solution
Alcoholic potassium hydroxide solution Oxalic acid solution Sodium carbonate solution
Ammonium iron (II) sulfate solution Perchloric acid Sodium chloride solution
Ammonium thiocyanate solution Potassium bromate solution Sodium hydoxide solution
Barium chloride solution Potassium bromate bromide solution Sodium nitrite solution
Bromide bromate solution Potassium dichromate solution Sodium thiosulfate solution
Cerium (IV) sulfate solution Potassium hydroxide solution Sulfuric acid
EDTA solution Potassium iodate solution Tetra-n-butylammonium hydroxide solution
Hydrochloric acid Potassium permanganate solution Zinc sulfate solution
Iodine solution Potassium thiocyanate solution  
Iron (II) sulfate solution Silver nitrate solution  

Limitations of use / Operating limits

Chlorinated and fluorinated hydrocarbons or chemical combinations which form deposits may make the piston difficult to move or may cause jamming.

Compatibility of the instrument for a special application (e.g., trace material analysis) must be checked by the user. For additional information, please contact the manufacturer.

The instrument is not autoclavable!

This instrument is designed for titrating liquids, observing the following physical limits:

  • +15 °C to +40 °C of instrument and reagent
  • Vapor pressure up to 500 mbar
  • Viscosity up to 500 mm²/s
  • Altitude: maximum 3000 m above sea level
  • Relative humidity: 20% to 90%.