ISO 9001 and GLP guidelines require routine calibration (approx. every 3-12 months) and readjustment of measuring instruments if necessary. This frequently time consuming task can be finished in seconds with BRAND Liquid Handling Instruments.

 

1.

Open housing by sliding the latch and removing the front.

2.

Lift gear lock lever to release.

3.

Pull the red knob and set the display to actual delivered volume. Reposition red knob and gear lock lever to the original position.

4.

Replace housing – done! Alteration of factory setting is indicated by a red recalibration flag.