Transferpette® -8/-12 electronic
Precision
Precision values for the Transferpette®-8/-12 electronic
Volume range
µl
Partial volume
µl
A*
< =+ %
CV*
< =
%
Substeps
µl
Tip type
µl
0.5  -  10  10 1.2 0.8 0.01 20
   5 2.0 1.5    
   1 8.0 4.0    
1  -   20  20 1.0 0.5 0.02 20
  10 2.0 1.0    
  2 8.0 3.0    
5  -   100  100 0.8 0.25 0.1 200/300
   50 1.6 0.4    
   10 4.0 1.5    
10  -  200  200 0.8 0.25 0.2 200/300
  100 1.4 0.4    
   20 4.0 1.3    
15  -  300 300 0.6 0.25 0.5 300
  150 1.2 0.4    
  30 3.0 1.2    

A* = Accuracy,   CV* = Coefficient of variation

 

Final test values related to the nominal capacity (maximum volume) indicated on the instrument, obtained when instrument and distilled water are equilibrated at ambient temperature (20 °C). According to DIN EN ISO 8655.