barrierefrei



Home : Products : Liquid Handling : Microliter pipettes : Transferpette® -8/-12 electronic : Precision


Precision values for the Transferpette®-8/-12 electronic
Volume range
µl
Partial volume
µl
A*
< =+ %
CV*
< =
%
Substeps
µl
Tip type
µl
0.5  -  10  10 1.2 0.8 0.01 20
   5 2.0 1.5    
   1 8.0 4.0    
1  -   20  20 1.0 0.5 0.02 20
  10 2.0 1.0    
  2 8.0 3.0    
5  -   100  100 0.8 0.25 0.1 200/300
   50 1.6 0.4    
   10 4.0 1.5    
10  -  200  200 0.8 0.25 0.2 200/300
  100 1.4 0.4    
   20 4.0 1.3    
15  -  300 300 0.6 0.25 0.5 300
  150 1.2 0.4    
  30 3.0 1.2    

A* = Accuracy,   CV* = Coefficient of variation

 

Final test values related to the nominal capacity (maximum volume) indicated on the instrument, obtained when instrument and distilled water are equilibrated at ambient temperature (20 °C). According to DIN EN ISO 8655.




Titel: Precision

Link: http://www.brand.de/en/products/liquid-handling/microliter-pipettes/motor-diven-multichannel-pipette-transferpetter-8-12-electronic/precision/printer.html


Copyright © www.brand.de 2012
All rights reserved.
Copyright contents may not be reproduced in any manner. Pictures and texts are not binding and subject to changes. We have carefully checked the contents of this website. However, BRAND GMBH + CO KG cannot accept responsibility for errors, omissions or possible obsolescence.

Close window