| Precision values for the Transferpette® electronic | |||||
|
Volume range µl |
Partial volume µl |
A* < =+ % |
CV* < = % |
Substeps µl |
Tip type µl |
| 0,5 - 10 | 10 | 1.0 | 0.4 | 0.01 | 20 |
| 5 | 1.5 | 0.8 | |||
| 1 | 5.0 | 2.0 | |||
| 2 - 20 | 20 | 1.0 | 0.4 | 0.02 | 20 |
| 10 | 1.5 | 0.8 | |||
| 2 | 5.0 | 2.5 | |||
| 20 - 200 | 200 | 0.8 | 0.2 | 0.2 | 200 |
| 100 | 1.2 | 0.3 | |||
| 20 | 4.0 | 0.6 | |||
| 100 - 1000 | 1000 | 0.6 | 0.2 | 1.0 | 1000 |
| 500 | 1.0 | 0.3 | |||
| 100 | 3.0 | 0.6 | |||
| 500 - 5000 | 5000 | 0.6 | 0.2 | 5.0 | 5000 |
| 2500 | 1.0 | 0.3 | |||
| 500 | 3.0 | 0.6 | |||
|
A* = Accuracy, CV* = Coefficient of variation |
|||||
Final test values related to the nominal capacity (maximum volume) indicated on the instrument, obtained when instrument and distilled water are equilibrated at ambient temperature (20 °C). According to DIN EN ISO 8655.