Transferpette® -8/-12
Easy Calibration
The factory calibration of the Transferpette® is adjusted to aqueous solutions at 20°C (68°F). For pipetting at substantially different temperatures or viscosities, the Transferpette® can be easily and quickly recalibrated by the user.

Example: Gravimetric testing yields a delivered volume of 98.5 µl at a set volume of 100 µl (with a medium which has a different viscosity than water, or in a cold room). The instrument can be calibrated (adjusted) in five easy steps without tiny set screws.

 

1.  Pull off tip ejector cap.

2.  Remove securing plate.

3. Set volume to actual value
   (e.g., 98.5 μl).

4. Replace securing plate.

5. Replace tip ejector cap. All done!