Accuracy table
HandyStep® electronic repetitve pipette with PD-Tips from BRAND, conformity certified
| HandyStep® electronic with PD-Tip | Volume range | Subdivision | Nominal volume (A* ≤ ± %) | Nominal volume (CV* ≤ %) | |||||||
| 100 % | 50 % | 10 % | 1 % | 100 % | 50 % | 10 % | 1 % | ||||
| 0.1 ml | 1 µl - 100 µl | 1 µl - 100 µl | 0.1 µl | 1.0 | 1.2 | 1.6 | 16 | 0.5 | 1.0 | 2.0 | 12 |
| 0.5 ml | 5 µl - 500 µl | 5 µl - 100 µl 100 µl - 500 µl |
0.1 µl 1 µl |
0.9 | 0.9 | 0.9 | 9 | 0.25 | 0.5 | 1 | 6 |
| 1.0 ml | 1 µl - 100 µl | 10 µl - 1 µl | 1 µl | 0.6 | 0.6 | 0.9 | 8 | 0.2 | 0.3 | 0.6 | 4 |
| 1.25 ml | 12.5 µl - 1250 µl | 12.5 µl -
100 µl 100 µl - 1000 µl 1 ml - 1.25 ml |
0.5 µl 1 µl 10 µl |
0.6 | 0.9 | 0.9 | 8 | 0.15 | 0.3 | 0.6 | 3.5 |
| 2.5 ml | 25 µl - 2500 µl | 25 µl - 1000 µl 1 ml - 2.5 ml |
1 µl 10 µl |
0.5 | 0.8 | 0.8 | 8 | 0.1 | 0.2 | 0.4 | 2.5 |
| 5.0 ml | 50 µl - 5000 µl | 50 µl -
1000 µl 1 ml - 5 ml |
1 µl 10 µl |
0.5 | 0.8 | 0.8 | 8 | 0.08 | 0.15 | 0.3 | 1.5 |
| 10.0 ml | 1 µl - 100 µl | 100 µl - 10 ml | 10 µl | 0.4 | 0.4 | 0.5 | 5 | 0.08 | 0.15 | 0.25 | 1.25 |
| 12.5 ml | 125 µl - 12.5 ml | 125 µl - 1000 µl 1 ml - 10 ml 10 ml - 12.5 ml |
5 µl 10 µl 100 µl |
0.4 | 0.5 | 0.5 | 5 | 0.08 | 0.15 | 0.25 | 1.25 |
| 25.0 ml | 250 µl - 25 ml | 250 µl - 10
ml 10 ml - 25 ml |
10 µl 100 µl |
0.3 | 0.3 | 0.3 | 3 | 0.08 | 0.15 | 0.25 | 1.25 |
| 50.0 ml | 500 µl - 50 ml | 500 µl - 10 ml 10 ml - 50 ml |
10 µl 100 µl |
0.3 | 0.3 | 0.3 | 3 | 0.08 | 0.15 | 0.25 | 1.25 |
| Error limits refer to the nominal volumes and partial volumes relative to the PD-Tip, obtained with instrument and distilled water at equilibrium with ambient temperature at 20 °C, and with smooth operation. The error limits defined in ISO 8655 are not exceeded. A = Accuracy, CV = Coefficient of variation |


