NEW! Synthetic coating for better protection
Volumetric flasks, BLAUBRAND®, PURprotect, Class A
The PUR coating envelops the volumetric flasks like a protective skin.
In the event of glass breakage, the safety coating helps contain the glass fragments. Compared to uncoated glass flasks the electrostatic charge is not increased. The light blue coating facilitates optical distinction.
The maximum operating temperature at dry heat is 135°C (exposure time
< 30 min). Frequent autoclaving at 121 °C reduces splintering protection. Cleaning up to max. 95 °C.
>> Watch PURprotect in a short flash movie!!
NEW! Testing flasks for dispensers
Volumetric flask with 3 marks, DKD-calibrated, BLAUBRAND®,
class A, conformity certified
DURAN®, DIN EN ISO 1042.
Calibrated to contain (TC, In). Incl. DKD-certificate. The volumetric flask
with 3 marks is used to check the functioning of a dispenser. The mark
in the middle shows the nominal volume, the upper and lower marks show
the error limits as specified in the table below. If the error limit is excee-
ded even with repeated measurements, the dispenser is defective.
The DKD-calibrated volumetric flask does not replace the gravimetric test
of the monitoring of measuring instruments according to ISO 8655.
|