| Précision de la Transferpette®-8/-12 electronic | |||||
|
Gamma de volume µl |
Volume de la fraction µl |
E* < =+ % |
CV* < = % |
Pas intermediaires µl |
Type de pointe µl |
| 0.5 - 10 | 10 | 1.2 | 0.8 | 0.01 | 20 |
| 5 | 2.0 | 1.5 | |||
| 1 | 8.0 | 4.0 | |||
| 1 - 20 | 20 | 1.0 | 0.5 | 0.02 | 20 |
| 10 | 2.0 | 1.0 | |||
| 2 | 8.0 | 3.0 | |||
| 5 - 100 | 100 | 0.8 | 0.25 | 0.1 | 200/300 |
| 50 | 1.6 | 0.4 | |||
| 10 | 4.0 | 1.5 | |||
| 10 - 200 | 200 | 0.8 | 0.25 | 0.2 | 200/300 |
| 100 | 1.4 | 0.4 | |||
| 20 | 4.0 | 1.3 | |||
| 15 - 300 | 300 | 0.6 | 0.25 | 0.5 | 300 |
| 150 | 1.2 | 0.4 | |||
| 30 | 3.0 | 1.2 | |||
|
E* = Exactitude, CV* = Coefficient de variation |
|||||
Les valeurs finales d‘essai se rapportent au volume nominal imprimé sur l‘appareil (= volume maxi), conformément aux exigences de la norme DIN EN ISO 8655.